  • Instrumentos de Calibración de Termopares de 0.1 hasta 0.01, °F e °C
    Tipos J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), (T-DIN) e P (Platinel): -13.000 hasta 80.000 Mv
  • Instrumentos de Calibración de RTDs de 0.1 hasta 0.01, °F e °C
    Pt 10, 50, 100, 200, 500 and 1000O (3850) Pt 100O (alfa = 3902, 3916, 3926) Cu 10 e 50O e Ni 120º
  • Calibre con confianza
    Resolución 0.01, ° F y ° C; Precisión de 0,2 ° Sensor de Precisión de junta fría con una precisión de 0,05 ° C Sensor de junta fría en bloque isotérmico con trazabilidad NIST
  • Troubleshoot RTD Sensors and Instruments
  • Sensores y Instrumentos de Solución de Problemas RTD
    Detección automática de 2, 3 o 4 conexiones de cables indica cuando se abren conexiones de hilos individuales Revise el Instrumento Adecuado mediante la medición de la corriente de excitación
  • Calibraciones Rápidas y Fáciles
    Selección instantánea de 3 salidas con el interruptor EZ-CHECK Calibraciones Rápidas y Fáciles
    Selección instantánea de 3 salidas con el interruptor EZ-CHECK Encuentre fácilmente los puntos de viaje con el EZ-DIAL Knob de dos velocidades y desplazamiento automático
  • Calibraciones Rápidas y Fáciles
    Selección instantánea de 3 salidas con el interruptor EZ-CHECK Compruebe la Resistencia, Temperatura de Unión fría y Salidas
    Pantalla secundaria de resistencia para RTDs y Configuración de la compensación de milivoltios y de la unión fría de temperatura para los termopares
  • Compruebe la Resistencia, Temperatura de Unión fría y Salidas
    Utilice en la Fábrica, Campo o Tienda
    Incluye Galoshes y Cables de prueba MT Interruptores de Perfil Bajo resistentes a daños accidentales Tornillos de bronce para los cables termopar, más enchufe de termopar en miniatura
  • Realizar Tratamiento Térmico en Estudios y Pruebas de Precisión
    Cumple o supera los requisitos de la norma AMS 2750 tanto como un instrumento secundario estándar cómo un instrumento de prueba de campo

Descripción

CONNECCIONES TERMOPARES
La simulación o lectura de termopares requiere el uso de termopares o cables de extensión para termopar. Conecte los cables de termopares en los enchufes miniatura o coloque los cables de termopares desnudos en el bloque de latón debajo de los tornillos.
El CL543B tiene dos enchufes tipo plátano (1+ y 2-) montados en el extremo superior de la carcasa. Estos no son de temperatura compensada y se deben utilizar solamente para las señales en milivoltios.

CONNECCIONES RTD
La simulación o la lectura de RTDs requiere el uso de hilo de cobre. Ligue 2, 3 o 4 cables en los enchufes correspondientes en el calibrador. Para una fuente de RTD, el CL543B simula el RTD (+) de los enchufes 1 e 4 y el RTD (-) de los enchufes 2 y 3. Durante la lectura de un sensor RTD, el CL543B utiliza un circuito patentado para detectar automáticamente si 2, 3 o 4 cables están conectados. Esto es útil para la solución de problemas de conexiones del sensor.

Specifications
Operating Temperature Range: -20 to 60°C (-5 to 140°F)
Storage Temperature Range: -30 to 60°C (-22 to 140°F)
Temperature Effect: ≤±25 ppm/°C of range
Cold Junction Sensor: ≤±0.01% °/° from ambient temperature (°C or °F)
Relative Humidity Range:
    10% ≤RH ≤90% (0 to 35°C), non-condensing
    10% ≤RH ≤70% (35 to 60°C), non-condensing
Normal Mode Rejection: 50/60 Hz, 50 dB
Common Mode Rejection: 50/60 Hz, 120 dB
Dimensions: 143 L x 76 W x 41 mm H (5.63 x 3 x 1.60")
Weight: 0.34 kg (12.1 oz) with boot and batteries
Batteries: Four “AA” alkaline 1.5V (LR6) (included), optional NiMh rechargeable battery kit [120 Vac (North America only); charger, 4 NiMh batteries, AC and DC cords
Battery Life: 50 hours, low battery indication with nominal 1 hour of operation left
Over-Voltage Protection: Up to 60 vrms (rated for 30 seconds), protection against misconnection
Alerts: Red LED displays OVERLOAD or out of range conditions
Display: High contrast graphic liquid crystal display, LED backlighting for use in low lit areas
Voltage Source
Ranges and Resolution: -13.000 to 80.000 mV and -13.0000 to 80.0000 mV
Accuracy: ±(0.008% of setting + 0.006 mV)
Source Current: ≥10 mA
Output Impedance: <0.3Ω
RMS Noise: ≤±0.0005 mV from 0.1 to 10 Hz
Short Circuit Duration: Infinite Voltage Read
Ranges and Resolution: -13.000 to 80.000 mV and -13.0000 to 80.0000 mV
Accuracy: ±(0.008% of reading + 0.006 mV)
Input resistance: ≥10 MΩ Thermocouple Source
Accuracy: ±(0.008% of setting + 0.006 mV)
Cold Junction Compensation: ±0.05°C (±0.09°F); thermistor traceable to NIST for 11 years
Output Impedance: <0.3Ω
Source Current: >10 mA (drives 80 mV into 10Ω)
RMS Noise: ≤±0.0005 mV from 0.1 to 10 Hz
Thermocouple Read
Accuracy: ±(0.008% of reading + 0.006 mV)
Cold Junction Compensation: ±0.09°F (±0.05°C); thermistor traceable to NIST for 11 years
Input Impedance: >10 MΩ
Open Thermocouple Threshold: 10,000Ω nominal
Pulse: <10 microamp pulse for 300 milliseconds
RTD and O Source
3- and 4-Wire Accuracy:
    From 1 to 10.2 mA External Excitation Current: ±(0.015% of setting + 0.05Ω)
    Below 1 mA of External Excitation Current: Add (0.025 mV/mA excitation current)
2-Wire Accuracy: Add 0.1Ω to 3-wire and 4-wire accuracy
Resistance Ranges:
    400Ω Range: 0.00 to 401.00 and 0.000 to 401.000
    4000Ω Range: 0.0 to 4010.0 and 0.00 to 4010.00
RMS Noise:
    400Ω Range: ≤±0.005Ω from 0.1 to 10 Hz
    4000Ω Range: ≤±0.05Ω from 0.1 to 10 Hz
Allowable Excitation Current Range:
    400Ω Range: 10.2 mA max; steady or pulsed/intermittent
    4000Ω Range: 1 mA max; steady or pulsed/intermittent
Pulsed Excitation Current Compatibility: DC to 0.01 second pulse width, 0.025 mV, mA excitation current
RTD and Ohms Read
Resistance Ranges:
    400Ω Range: 0.00 to 401.00 and 0.000 to 401.000
    4000Ω Range: 0.0 to 4010.0 and 0.00 to 4010.00
Accuracy: ±(0.015% of reading + 0.05Ω)
Excitation Current: 0.9 mA to 401Ω, 0.4 mA to 4010Ω (nominal)
Hacer Pedido(Especifique el Número de Modelo)
Número de parte/Desc.
Cant.
Descripción
Quick Ship
CL543B
USD$1,668.00
Disponible en: Disponible
 
Thermocouple and RTD temperature calibrator
Todos los productos en este sitio se muestran en la moneda de EE.UU.
NOTA: Comes complete with NIST certificate, 4 “AA” alkaline batteries, certificate of calibration card, rubber boot, deluxe hands free carrying case, mA/V test leads, 1 red and 1 black lead with banana plugs and alligator clips, RTD wire kit, 2 red and 2 black leads with banana plugs and spade lugs, and operator’s manual.